Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Abbott Medical recalls Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from

Class IIIofficial

FDA Class III

FDA classified this recall as Class III: the product is unlikely to cause health problems but breaks the agency's rules (a labeling mistake, for example).

Hazard
Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.
Products
  • Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from
Where it was sold

Worldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, ILLINOIS, INDIANA, KENTUCKY, MICHIGAN, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEW JERSEY, NEW YORK, NORTH CAROLINA, OHIO, OREGON, PENNSYLVANIA, SOUTH CAROLINA, TENNESSEE, TEXAS, VIRGINIA, WASHINGTON and WISCONSIN. The country of Australia.

States: Alaska, Alabama, Arkansas, California, Colorado, Connecticut, District of Columbia, Florida, Georgia, Illinois, Indiana, Kentucky, Michigan, Missouri, Mississippi, Montana, North Carolina, Nebraska, New Jersey, New York, Ohio, Oregon, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Washington, Wisconsin

Units affected
130 transmitters
About this recall
Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from the patient s implanted heart device to Merlin.net" Patient Care Network (PCN) to facilitate remote monitoring by using inductive (EX1100 Model) or RF telemetry (EX1150 Model) to read device data and transmitting to the Merlin.net PCN server.