MEDLINE INDUSTRIES, LP - Northfield recalls Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
- Products
-
- Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter
- Where it was sold
-
US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.
Nationwide
- Barcodes
100801960327611008019603287710080196503391100801965034071008019650338410080196032914- Units affected
- 620,735 kits
- About this recall
- Tracheostomy care kits: (1) Medline Tracheostomy Clean and Care Tray with 14 Fr Suction Catheter, REF DYND40580, Sterile; (2) Medline Tracheostomy Clean and Care Tray, REF DYND4061030, Sterile; (3) Medline Basic Tracheostomy Care Kit, REF CC3T4691A, Sterile; (4) Medline Tracheostomy Care Kit with Glove, REF CC4681A, Sterile; (5) Medline Mini Tracheostomy Care Kit, REF CC3T3030A, Sterile; and (6) Medline Mini Tracheostomy Care Kit, REF DYND40622, Non-sterile.