MEDLINE INDUSTRIES, LP - Northfield recalls (1) Nurse Kit
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
- Products
-
- (1) Nurse Kit
- Where it was sold
-
US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.
Nationwide
- Barcodes
4019532719664740195327153695- Units affected
- 550 kits
- About this recall
- (1) Nurse Kit, SKU #EDUC1025, Case UPC 40195327196647, Non-sterile, For Training Purposes Only, Not For Human Use.; and (2) Nursing Tote Kit, SKU #EDUC5056A, Case UPC 40195327153695, Non-sterile, For Training Purposes Only, Not For Human Use.