LivaNova Deutschland GmbH recalls Essenz HLM (Heart-Lung Machine)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.
- Products
-
- Essenz HLM (Heart-Lung Machine)
- Company
- LivaNova Deutschland GmbH
- Where it was sold
-
US Nationwide distribution in the states of CA, VA, GA, IN, NJ, FL, CO, MS, NV.
States: California, Colorado, Florida, Georgia, Indiana, Mississippi, New Jersey, Nevada, Virginia
- Barcodes
04033817903062- Units affected
- 38
- About this recall
- Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software