Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

LivaNova Deutschland GmbH recalls Essenz HLM (Heart-Lung Machine)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.
Products
  • Essenz HLM (Heart-Lung Machine)
Company
LivaNova Deutschland GmbH
Where it was sold

US Nationwide distribution in the states of CA, VA, GA, IN, NJ, FL, CO, MS, NV.

States: California, Colorado, Florida, Georgia, Indiana, Mississippi, New Jersey, Nevada, Virginia

Barcodes
04033817903062
Units affected
38
About this recall
Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software