Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Orthoscan, Inc. recalls Ziehm Solo. Digital Mobile C-arm

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.
Products
  • Ziehm Solo. Digital Mobile C-arm
Where it was sold

U.S. (including Puerto Rico).

States: Puerto Rico

Units affected
28
About this recall
Ziehm Solo. Digital Mobile C-arm