Orthoscan, Inc. recalls Ziehm Vision system
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.
- Products
-
- Ziehm Vision system
- Company
- Orthoscan, Inc.
- Where it was sold
-
U.S. (including Puerto Rico).
States: Puerto Rico
- Units affected
- 3
- About this recall
- Ziehm Vision system, x-ray, tomography, computed.