Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Mobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
Products
  • MICRO-X Rover Mobile X-ray System
Where it was sold

US: TX, PR, MD, CA

States: California, Maryland, Puerto Rico, Texas

Barcodes
9357123000037
Units affected
6
About this recall
MICRO-X Rover Mobile X-ray System, # MXU-RV35