Micro-X Ltd. recalls
Micro-X Ltd. is named on 7 recalls in Recall Relay's records since November 2023, from FDA (9 recalled products: FDA lists each product as its own recall); the most recent was March 27, 2026.
All 9 recalled products, newest first.
-
Class IIofficial
Micro-X Ltd. recalls Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71X-Ray generator may malfunction resulting in the X-Ray being inoperable.
-
Class IIofficial
Micro-X Ltd. recalls Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71Potential for internal fastening mechanism within generator of mobile x-Ray to fail.
-
Class IIofficial
Micro-X Ltd. recalls Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number:MXU-RV35 Product Description: Rover MPotential for internal fastening mechanism within generator of mobile x-Ray to fail.
-
Class IIofficial
Micro-X Ltd. recalls Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: RoverPotential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
-
Class IIofficial
Micro-X Ltd. recalls Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV35 Product Description: RoverPotential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
-
Class IIofficial
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray SystemDuring internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.
-
Class IIofficial
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray SystemMobile x-ray system receiving/storage(operation) requirements to be followed: Temp:14 to 131(50 to 86) degrees F, RH:10% to 86%(30% to 60%) allow condensation dry time, Atmospheric Pressure:50kPa to 106 kPa(70kPa to 106kPa), Altitude:-102ft to 12,000 ft, (Maximum Gradient:9 degrees Fahrenheit), which mitigate risk of capacitor malfunction, unpredictable behavior, overheating, unintended start-up.
-
Class IIofficial
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray SystemMobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
-
Class IIofficial
Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray SystemShots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.