Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medtronic Neuromodulation recalls Medtronic SynchroMed II

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.
Products
  • Medtronic SynchroMed II
Where it was sold

Worldwide

Nationwide

Barcodes
00613994483195006431691008240064316934521800643169345225006431693452320064316938417000643169384187006431693841940064316938420000643169384217006431693842240064316938423100643169508156006431695301260064316963051200643169701007006431697010140064316970102100643169701038006431697010450064316970105200643169701069006431697010760064316970108300643169701090006431697323460064316973235300643169732360006431697323770064316973238400643169732391006431697324070064316973241400643169732421006431697324380064316998422600643169999930006431699999470064316999995400643169999961006431699999780064316999998500763000000004007630000000110076300000002800763000000035007630000187400076300001876400763000080747007630001227510076300012276800763000122775007630001227820076300012279900763000122805007630001228120076300012282900763000122836007630001228500076300042182300763000421830007630004218470076300042185400763000421861007630004218780076300042188500763000421892007630004219080076300042192200763000422615007630006042020076300063410000763000689476007630006894830076300068949000763000689506007630006895130076300068952000763000689537007630006895510076300068956800763000689575
Units affected
78532 units
About this recall
Medtronic SynchroMed II, Model 8637-40, Programmable pump