Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
Products
  • MICRO-X Rover Mobile X-ray System
Where it was sold

Worldwide and US Nationwide Distribution: NJ, PR, CA

Nationwide

Barcodes
9357123000037
Units affected
5
About this recall
MICRO-X Rover Mobile X-ray System, # MXU-RV35