Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.
Products
  • MICRO-X Rover Mobile X-ray System
Where it was sold

US Nationwide Distribution: TX, CA, OH, ID, WA, AZ

Nationwide

Barcodes
09357123000013
Units affected
23
About this recall
MICRO-X Rover Mobile X-ray System, # MXU-RV19