Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

MEDLINE INDUSTRIES, LP - Northfield recalls HUDSON RCI mBrace

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Complaint defects have been noted regarding the adhesive elements of the device and the frame of the holder putting additional pressure on the patient's cheekbones and upper lip which may cause dermal injury depending on the device positioning.
Products
  • HUDSON RCI mBrace
Where it was sold

US and Canada

Barcodes
20193489005339
Units affected
768 units
About this recall
HUDSON RCI mBrace, ET Tube Holder with Bite Block and 4 Point Head Strap, REF DYNJMBRC4B; tracheal tube fixation device