Maquet Cardiovascular, LLC recalls Product REF 4050-100N
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Express chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
- Products
-
- Product REF 4050-100N
- Company
- Maquet Cardiovascular, LLC
- Where it was sold
-
Worldwide distribution: US (nationwide) and OUS (Foreign) to countries of: Australia, Brazil, Canada, Luxembourg, Netherlands, Norway, Spain and UK
Nationwide
- Barcodes
00650862115147- Units affected
- (7,501 cases/45,006 devices US); no OUS
- About this recall
- Product REF 4050-100N, DRAIN, EXPRESS BRU W/AC, Blood Recovery Unit