Maquet Cardiovascular, LLC recalls
Maquet Cardiovascular, LLC is named on 13 recalls in Recall Relay's records since November 2023, from FDA (22 recalled products: FDA lists each product as its own recall); the most recent was June 2, 2026.
All 22 recalled products, newest first.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterileGetinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Heartstring III Proximal Seal SystemThree failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Heartstring III Proximal Seal SystemThree failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Heartstring III Proximal Seal System. Intravascular anastomosis occluder.Three failure modes have been identified: failure of the Heartstring Seal to load, failure of the Heartstring Seal to deploy into the aortotomy, and failure of the deployed Heartstring Seal to provide adequate hemostasis.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 EndoscopicReports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm exThere have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
-
Class Iofficial
Maquet Cardiovascular, LLC recalls Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for uPotential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool
-
Class Iofficial
Maquet Cardiovascular, LLC recalls VasoView HemoPro 2 Endoscopic Vessel Harvesting System. Model Numbers: VH-4000. Indicated for use in minimallyPotential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Acrobat-i Positioner. Model Number C-XP-5000ZDeviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Acrobat SUV Vacuum Stabilizer SystemDeviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000ZDeviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
-
Class Iofficial
Maquet Cardiovascular, LLC recalls VasoView HemoPro Endoscopic Vessel Harvesting SystemThere were reports of the silicone detaching from the Jaws of the Harvesting Tool during use.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm ExThere were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm ExtThere were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls The VASOVIEW HemoPro Endoscopic Vessel Harvesting Systems are designed for use in conjunction with the 7 mm ExPotential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls The VASOVIEW HemoPro 2 Endoscopic Vessel Harvesting SystemPotential for there to be a fluid ingress into the handle of the harvesting tool, which may lead to the device not providing cautery or providing cautery when not intended.
-
Class Iofficial
Maquet Cardiovascular, LLC recalls Product REF 4000-100NExpress chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
-
Class Iofficial
Maquet Cardiovascular, LLC recalls Product REF 4050-100NExpress chest drains are pre-packaged with sterile water syringes which have bene recalled. If the device is not kept in an upright position, the patient may be exposed to infectious material. Subsequent hazards include but are not limited to infection, abscess, sepsis, and death.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Vasoview Hemopro Endoscopic Vessel Harvesting SystemRisk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Vasoview Hemopro Endoscopic Vessel Harvesting SystemRisk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Acrobat-i Vacuum Positioner SystemThe Positioner Arm may not tighten or lock during normal use.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.