Olympus Corporation of the Americas recalls Soltive SuperPulsed Laser System TFL FIBER CLEAVER
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.
- Products
-
- Soltive SuperPulsed Laser System TFL FIBER CLEAVER
- Where it was sold
-
US Nationwide. Global Distribution
Nationwide
- Units affected
- 236 units
- About this recall
- Soltive SuperPulsed Laser System TFL FIBER CLEAVER, Model Number TFL-AFC