FDAMedical devicesRecalled Status: Ongoing

Philips Medical Systems Nederland B.v. recalls Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
Products
  • Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
Where it was sold

Domestic: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MD, MN, MO, NM, NY, OH, OK, PA, RI, TX, UT, VA, WA, WI, WV, WY; International: Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Iraq, Italy, Japan, Netherlands, Norway, Philippines, Poland, Romania, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Vietnam;

States: Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Illinois, Indiana, Kansas, Kentucky, Maryland, Minnesota, Missouri, New Mexico, New York, Ohio, Oklahoma, Pennsylvania, Rhode Island, Texas, Utah, Virginia, Washington, Wisconsin, West Virginia, Wyoming

Units affected
172 units
About this recall
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;