Cook Medical Llc recalls MINC+ Benchtop Incubator
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
- Products
-
- MINC+ Benchtop Incubator
- Company
- Cook Medical Llc
- Where it was sold
-
Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM
States: California, Indiana, Louisiana, Maryland, New York
- Barcodes
00827002444294- Units affected
- 14 US; 101 OUS
- About this recall
- MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429