Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Enterix, Inc. recalls InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Numbe

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
InSure ONE failed sensitivity testing when used with Test Card Lot T241102, which may result in false negative results.
Products
  • InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Numbe
Company
Enterix, Inc.
Where it was sold

Worldwide - US Nationwide distribution in the states of California, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Nevada, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Texas, Washington, West Virginia and the country of Philippines.

States: California, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, North Carolina, New Jersey, Nevada, New York, Ohio, Pennsylvania, Texas, Virginia, Washington, West Virginia

Barcodes
108500930033132085009300331000850093003279
Units affected
201,426 test kits
About this recall
InSure ONE Fecal Immunochemical Test. Labeled as follows: 1. Insure ONE POL Patient Kit - 10 pack. Model Number: 90010. 2. Insure ONE POL Patient Kit - 25 Pack. Model Number: 90025. 3. InSure ONE POL Combo Pack. Model Number: 90030. InSure ONE is a fecal immunochemical test (FIT) that qualitatively detects human hemoglobin from blood in fecal samples.