Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Greiner Bio-One North America, Inc. recalls VACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium Oxalate

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Tubes have missing additive/anticoagulant resulting in clotting in most cases. A completely clotted sample would lead to no results being reported, which could lead to a delay in treatment or diagnosis. An erroneous glucose or lactose result from a partially clotted sample could lead to inappropriate treatment.
Products
  • VACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium Oxalate
Where it was sold

US Nationwide distribution in the states of CT, FL, IA, IL, MO, NC, NY, PA, TN, and VA.

States: Connecticut, Florida, Iowa, Illinois, Missouri, North Carolina, New York, Pennsylvania, Tennessee, Virginia

Units affected
495,600 tubes
About this recall
VACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium Oxalate, Item #454297, 13x75 grey cap-black ring, non-ridged.