Greiner Bio-One North America, Inc. recalls VACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium Oxalate
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Tubes have missing additive/anticoagulant resulting in clotting in most cases. A completely clotted sample would lead to no results being reported, which could lead to a delay in treatment or diagnosis. An erroneous glucose or lactose result from a partially clotted sample could lead to inappropriate treatment.
- Products
-
- VACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium Oxalate
- Where it was sold
-
US Nationwide distribution in the states of CT, FL, IA, IL, MO, NC, NY, PA, TN, and VA.
States: Connecticut, Florida, Iowa, Illinois, Missouri, North Carolina, New York, Pennsylvania, Tennessee, Virginia
- Units affected
- 495,600 tubes
- About this recall
- VACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium Oxalate, Item #454297, 13x75 grey cap-black ring, non-ridged.