Greiner Bio-One North America, Inc. recalls VACUETTE¿ TUBE 8 ml CAT Serum Separator Clot Activator 16x100 red cap-yellow ring
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The firm received reports that some blood collection tubes have experienced issues with clotting and poor separation due to missing additive. The failure of the device would likely require sample recollection which could result in a delay of treatment.
- Products
-
- VACUETTE¿ TUBE 8 ml CAT Serum Separator Clot Activator 16x100 red cap-yellow ring
- Where it was sold
-
Nationwide Distribution to states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV and WY.
Nationwide
- Units affected
- 1,008,000pcs (840 cartons)
- About this recall
- VACUETTE¿ TUBE 8 ml CAT Serum Separator Clot Activator 16x100 red cap-yellow ring, non-ridged