Olympus Corporation of the Americas recalls SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.
- Products
-
- SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Canada, Europe, Hong Kong, Australia, Japan, Singapore, and Latin America.
Nationwide
- Units affected
- 82,200 total units (5 per box)
- About this recall