Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Waldemar Link GmbH & Co. KG (Mfg Site) recalls Link Patella Glide Resection Guide- For the implantation of the Endo-Model Knee System as part of a total knee

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Mislabeled: Angle and height of the device marked as L(large) corresponds to the small/medium dimensions due to a product mix-up. This may result in too little bone material resected and not fit correctly leading to extended surgery time
Products
  • Link Patella Glide Resection Guide- For the implantation of the Endo-Model Knee System as part of a total knee
Where it was sold

US Nationwide distribution in the state of NH.

States: New Hampshire

Barcodes
04026575245253
Units affected
1 unit
About this recall
Link Patella Glide Resection Guide- For the implantation of the Endo-Model Knee System as part of a total knee arthroplasty procedure. Product Code: 15-2530/05