Randox Laboratories Ltd. recalls Calibration Serum Level 3 CAL2351

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.
Products
  • Calibration Serum Level 3 CAL2351
Where it was sold

US Nationwide and Puerto Rico.

Nationwide

Barcodes
05055273200966
Units affected
269 kits
About this recall
Calibration Serum Level 3 CAL2351