Randox Laboratories Ltd. recalls Calibration Serum Level 3 CAL2351
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.
- Products
-
- Calibration Serum Level 3 CAL2351
- Company
- Randox Laboratories Ltd.
- Where it was sold
-
US Nationwide and Puerto Rico.
Nationwide
- Barcodes
05055273200966- Units affected
- 269 kits
- About this recall
- Calibration Serum Level 3 CAL2351