Cardinal Health 200, LLC recalls Cardinal Health MEDI-VAC CRD Suction Canister Liner
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.
- Products
-
- Cardinal Health MEDI-VAC CRD Suction Canister Liner
- Company
- Cardinal Health 200, LLC
- Where it was sold
-
US, Canada, Australia and New Zealand
- Units affected
- 203244 units
- About this recall
- Cardinal Health MEDI-VAC CRD Suction Canister Liner, 1000 mL Suction Canister Liner Kit with 5 mm 1.8 m Tube, REF 65651-516