Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Moximed, Inc. recalls MISHA Knee System Implant Small

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.
Products
  • MISHA Knee System Implant Small
Company
Moximed, Inc.
Where it was sold

US Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA.

States: California, Colorado, New York, Ohio, Oregon, Texas, Virginia

Units affected
115
About this recall
MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.