Moximed, Inc. recalls MISHA Knee System Implant Small
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.
- Products
-
- MISHA Knee System Implant Small
- Company
- Moximed, Inc.
- Where it was sold
-
US Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA.
States: California, Colorado, New York, Ohio, Oregon, Texas, Virginia
- Units affected
- 115
- About this recall
- MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.