Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Accure Acne, Inc recalls Accure Laser System

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Laser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering.
Products
  • Accure Laser System
Company
Accure Acne, Inc
Where it was sold

US distribution to states of: CO, CA, CT, NY, NJ, MN, LA, OH, MA, TX

States: California, Colorado, Connecticut, Louisiana, Massachusetts, Minnesota, New Jersey, New York, Ohio, Texas

Barcodes
08059173392230
Units affected
22
About this recall
Accure Laser System, Model: PFMS00004