Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medtronic Neuromodulation recalls Model A710 Intellis Clinician Programmer Application

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medtronic has discovered an instance in which the Intellis A710 Clinician Programmer Application (CP App) could no longer establish a session with the Implantable Neurostimulator (INS) which resulted in the inability for the clinician to adjust stimulation parameters. In the event of this issue, the patient may receive inadequate or excessive therapy and experience non-optimal pain relief.
Products
  • Model A710 Intellis Clinician Programmer Application
Where it was sold

Germany

Barcodes
00763000520076
Units affected
1 tablet
About this recall
Model A710 Intellis Clinician Programmer Application, version 2.0.97