Hologic, Inc. recalls PANTHER FUSION System REF 9121010000 The Panther System is an integrated nucleic acid testing system which ful

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to potential false positive Flu B results when a sample was also SARS-CoV-2 positive.
Products
  • PANTHER FUSION System REF 9121010000 The Panther System is an integrated nucleic acid testing system which ful
Where it was sold

Worldwide - U.S. Nationwide distribution including in the states of AK, AR, CA, CO, CT, FL, IL, IN, MA, ME, MI, MN, MO, MT, NC, ND, NH, NJ, NM, NV, NY, OH, PA, Puerto Rico, TX, UT, VA, VI, WA, WI, and WV. The countries of Austria, Australia, Belgium, Dominican Republic, Switzerland, Germany, Denmark, Spain, France, Great Britain, Hong Kong, Ireland, Italy, Korea, Kuwait, The Netherlands, Norway, New Zealand, Portugal, Qatar, and Sweden.

Nationwide

Units affected
771 systems (of which 313 are potentially impacted)
About this recall
PANTHER FUSION System REF 9121010000 The Panther System is an integrated nucleic acid testing system which fully automates all steps necessary to perform Aptima assays from sample processing through amplification, detection, and data reduction. The Panther Fusion System is an integrated, fully automated, in vitro diagnostic (IVD) clinical multiplex test system. It is capable of performing nucleic acid based tests from sample processing through amplification, detection, and data reduction.