Hologic, Inc. recalls Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Potential weak pouch seal on Panther Fusion GBS assay cartridge, causing potential for invalid or incorrect results leading to a possible delay of treatment.
- Products
-
- Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.
- Company
- Hologic, Inc.
- Where it was sold
-
AR, CA, CT, FL, IN, KS, LA, MA, ME, MI, NC, NJ, NV, NY, PR, TX, UT, VT, WA, WV, and Canada.
States: Arkansas, California, Connecticut, Florida, Indiana, Kansas, Louisiana, Massachusetts, Maine, Michigan, North Carolina, New Jersey, Nevada, New York, Puerto Rico, Texas, Utah, Vermont, Washington, West Virginia
- Barcodes
15420045510890- Units affected
- 380
- About this recall
- Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.