Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Phadia Ab recalls EliA GBM Wells

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Complaints that specific samples produced false positive EliA GBM results. A positive signal was present when these samples were tested for coating-solution reactivity using EliA wells without antigen.
Products
  • EliA GBM Wells
Company
Phadia Ab
Where it was sold

US Nationwide.

Nationwide

Barcodes
07333066010670
Units affected
124,928 kits
About this recall
EliA GBM Wells, Material Number 14551401, for Phadia 250, 2500, and 5000. In vitro diagnostic