Elucent Medical Inc recalls EnVisio Patient Pad Transition Cushion
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There is a potential safety risk caused by physical disassociation of the back section from the bed frame of Skytron model series 3600, 6500, 6600 and 6700 operating room beds when used with the EnVisio Navigation System.
- Products
-
- EnVisio Patient Pad Transition Cushion
- Company
- Elucent Medical Inc
- Where it was sold
-
US Distribution: AZ, CA, CO, CT, FL, GA, IL, KY, MN, NC, ND, NJ, NY, OH, PA, SC, WI
States: Arizona, California, Colorado, Connecticut, Florida, Georgia, Illinois, Kentucky, Minnesota, North Carolina, North Dakota, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Wisconsin
- Units affected
- 283 (48 units)
- About this recall
- EnVisio Patient Pad Transition Cushion, Model/Catalog Number: ENVISIO-SYS Product Description: The EnVisio Patient Pad Transition Cushion is a component of the EnVisio Navigation System which is a medical device that detects the presence of the SmartClip Soft Tissue Marker(s) and provides real-time three-dimensional navigation during surgery.