Boston Scientific Corporation recalls ACCOLADE DR Pacemaker
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
- Products
-
- ACCOLADE DR Pacemaker
- Where it was sold
-
US (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands.
Nationwide
- Barcodes
00802526559174008025265592280080252655924200802526559273- Units affected
- 156,329 units (59,937 US, 96,792 OUS) in total
- About this recall
- ACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Number L331)