American Contract Systems, Inc. recalls Custom procedural convenience kits and trays
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
- Products
-
- Custom procedural convenience kits and trays
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Units affected
- 112 units
- About this recall
- Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) Breast Plastics Pack, kit number BBPL20D; b) MINOR PLASTIC, kit number WPMP16H