Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Boston Scientific Corporation recalls ESSENTIO DR Pacemaker

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
Products
  • ESSENTIO DR Pacemaker
Where it was sold

US (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands.

Nationwide

Barcodes
00802526558924008025265589620080252655898600802526559006
Units affected
156,329 units (59,937 US, 96,792 OUS) in total
About this recall
ESSENTIO DR Pacemaker, SL (Model Number L101, SL MRI (Model Number L111), EL (Model Number L121), and EL MRI (Model Number L131)