FDAMedical devicesRecalled Status: Ongoing

Boston Scientific Corporation recalls ALTRUA 2 DR Pacemaker

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
Products
  • ALTRUA 2 DR Pacemaker
Where it was sold

US (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands.

Nationwide

Barcodes
0080252655935800802526559372
Units affected
156,329 units (59,937 US, 96,792 OUS) in total
About this recall
ALTRUA 2 DR Pacemaker, SL (Model Number S702) and EL (Model Number S722)