Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

American Contract Systems, Inc. recalls Custom procedural convenience kits and trays

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Products
  • Custom procedural convenience kits and trays
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
0019107214411900191072170125001910721884580019107215510800191072173676
Units affected
1020 units
About this recall
Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L; b) OB/GYN DAVINCI PACK A RG BASIN, kit number HGOB98G; c) DAVINCI GYN PACK, kit number HHDV17; d) KIT, GYNE ROBOTIC, kit number MMGR44D; e) KIT, MAJOR LITHOTOMY, kit number MMJL16E; f) GYN PACK, kit number SMGY46; g) Lithotomy Pack, kit number THLH71C; h) ASC HYSTEROSCOPY PACK (PS 020634), kit number UIHY16AA; i) GYN LAPAROSCOPY HPES, kit number WELP67C; j) GYN LAPAROSCOPY HPES, kit number WELP67C-01; k) GYN LAPAROSCOPY HPES, kit number WELP67C-02