Nuwellis Inc recalls nuwellis AquaFlexFlow UF 500 Plus
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.
- Products
-
- nuwellis AquaFlexFlow UF 500 Plus
- Company
- Nuwellis Inc
- Where it was sold
-
US Nationwide Distribution and the country of Hong Kong
Nationwide
- Units affected
- 845 units
- About this recall
- nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System