Nuwellis Inc recalls nuwellis AquaFlexFlow UF 500 Plus

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
The AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.
Products
  • nuwellis AquaFlexFlow UF 500 Plus
Company
Nuwellis Inc
Where it was sold

US Nationwide Distribution and the country of Hong Kong

Nationwide

Units affected
845 units
About this recall