Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Covidien LP recalls Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPC

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
Products
  • Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPC
Where it was sold

Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Japan, Korea, Republic Of Kosovo, Mongolia, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.

Nationwide

Barcodes
1088452107172810884521071742108845210717662088452107176320884521071794
Units affected
46,812 US; 68,976 OUS
About this recall
Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT