Covidien is named on 26 recalls in Recall Relay's records since November 2023, from FDA (44 recalled products: FDA lists each product as its own recall); the most recent was June 24, 2026.
All 44 recalled products, newest first.
-
Class IIofficial
Covidien, LP recalls Dermalon 88861757-31 DERMALON* 4-0 BLU 75CM C13 88861757-31 88861799-31 DERMALON* 4-0 BLU 75CM C14 88861799-31Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
-
Class IIofficial
Covidien, LP recalls Surgipro SP-623 SURGIPRO* 2-0 BLU 75CM SC2 X36 SP-623Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
-
Class IIofficial
Covidien, LP recalls V-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436 VLOCMProduct may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
-
Class IIofficial
Covidien, LP recalls Surgipro II S8698GX SURGIPRO II BL 5/0 18 P-13 S8698GXProduct may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
-
Class IIofficial
Covidien, LP recalls V-Loc 180 Absorbable Wound Closure Device BOX BEDSIDE STAPLING IS KING BOX06614V4 BOX BY PASS FINAL BOX07405V1Product may contain a perforation or tear in the sterile barrier packaging. This issue may result in biocontamination of the sterile sutures, which may lead to infection, loss of vision if in ophthalmic utilization, or prolonged surgery.
-
Class IIofficial
Covidien, LLC recalls EndoFLIP System that includes: ENDOFLIP 1 SYSTEM EF-100 05391530810050 10884521809307 REFURBISHED SYSTEM RFG-EThe system's 24-hour usage limit results in catheters incorrectly flagged as expired, making them unavailable for clinical procedures.
-
Class IIofficial
Covidien LLC recalls Covidien Mon-a-Therm Esophageal Stethoscope with Temperature Sensor 400TM REF: 90042Temperature probe devices lack FDA clearance.
-
Class IIofficial
Covidien LLC recalls Covidien Mon-a-Therm General Purpose Temperature Probe REF: 90044Temperature probe devices lack FDA clearance.
-
Class IIofficial
Covidien LLC recalls Covidien Shiley Disposable Decannulation Plug REF DDCP UDI-DI code: 10884522004060 Product Codes that includeDue to product labeling incorrectly including the symbol indicating non-DEHP (Bis(2-ethylhexyl) phthalate as the product includes 30 to 35% DEHP.
-
Class IIofficial
Covidien LLC recalls Covidien Mon-a-Therm general purpose temperature probe size: 9 Fr REF: 90050Temperature probe devices lack FDA clearance.
-
Class IIofficial
Covidien, LP recalls Covidien EndoStitch" Polysorb Single Use Loading Unit; Product numbers: 170053 and 170071;The presence of holes in the pouch may compromise the package s ability to maintain the sterile barrier and protect the product from environmental exposure.
-
Class IIofficial
Covidien LLC recalls COVIDIEN Shiley Tracheostomy Tube Cuffed with Inner Cannula REF 6FEN UDI-DI code: 60884522007028 The devices aDue to customer complaint regarding incorrect display box labeling.
-
Class IIofficial
Covidien LLC recalls HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALDue to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.
-
Class IIofficial
Covidien, LP recalls Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
-
Class IIofficial
Covidien, LP recalls Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); BOX LOBECTOMIA VATS 3D; CFA manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
-
Class IIofficial
Covidien recalls Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100NBedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting
-
Class Iofficial
Covidien recalls Newport HT70 VentilatorVentilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.
-
Class Iofficial
Covidien recalls Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner CannulaTracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a delay to treatment and/or death.
-
Class IIofficial
Covidien LP recalls Chameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US)Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.
-
Class Iofficial
Covidien recalls McGRATH" MAC EMS Video LaryngoscopeBattery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.
-
Class IIofficial
Covidien recalls Covideien McGRATH MAC Video LaryngoscopeIFU update to address device's battery handling information.
-
Class Iofficial
Covidien recalls McGrath Mac Video LaryngoscopeBattery management system within Laryngoscope devices may deplete below the deign threshold which may result in thermal event followed by risk of explosion.
-
Class Iofficial
Covidien recalls Puritan Bennett 500 Series Ventilators (Description/REF): PURITAN BENNETT 560 VENTILATOR/4096600Any single affected ventilator should only be used for the 10- year labelled service life, but if longer ventilatory support is required using more than one ventilator over time, the use of these devices for more than 14 years of cumulative duration may pose harm to health due to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro released from some components in the ventilator gas pathway
-
Class IIofficial
Covidien, LLC recalls EsoFLIPDue to saline conductivity issues, their is a potential that dilation catheters may provide inaccurate esophageal diameter measurements that may lead to various harms.
-
Class IIofficial
Covidien, LP recalls Tri-Staple 2.0 Black Reinforced Intelligent Reload- Applications in abdominalPotential damage to the cartridge can result in poor staple formation and/or incomplete staple line, may cause a delay in surgery
-
Class IIofficial
Covidien LP recalls Palindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical TipProducts may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating
-
Class IIofficial
Covidien LP recalls Palindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical TipProducts may not be heparin coated as indicated on the label-leads to a potential loss of benefit due to the absence of the heparin coating
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Structural Balloon TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien recalls Situate Laparotomy SpongePacks including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-sterile, may lead to infection, tissue trauma, and sepsis, and if in the scannable area, would lead a "CLEAR" display.
-
Class IIofficial
Covidien, LP recalls Covidien Signia" Small Diameter Curved Tip Intelligent ReloadUnder certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
-
Class IIofficial
Covidien, LP recalls Signia" Small Diameter Curved Tip Intelligent ReloadUnder certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
-
Class IIofficial
Covidien, LP recalls Sofsilk Braided Silk suturesOne lot of product was exposed to more than the approved number of Ethylene Oxide (EO) sterilization cycles. This may decrease the tensile strength of the sutures over time, which could result in harms such as wound dehiscence, hemorrhage, blood loss, bleeding, tissue breakdown, peritonitis, unspecified infection, vision loss (when used in ophthalmic application), and/or prolonged surgery.
-
Class IIofficial
Covidien recalls Medtronic Invosmanufacturing defects to sensor cable connectors may result in error codes or non-functioning channel
-
Class IIofficial
Covidien recalls Medtronic Invosmanufacturing defects to sensor cable connectors may result in error codes or non-functioning channel
-
Class IIofficial
Covidien LP recalls Sofsilk Braided Silk sutures: CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 GS-831 SOFSILK* 4-0 BLK 75CM V20 X36 S-17Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
-
Class IIofficial
Covidien LP recalls Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCSpecific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
-
Class IIofficial
Covidien recalls Covidien Valleylab FT10Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable
-
Class IIofficial
Covidien LP recalls AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic mPackaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.