Philips Medical Systems Nederland B.v. recalls Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
Products
  • Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
Where it was sold

Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and TX. The country of Canada.

States: Colorado, Florida, Illinois, Mississippi, Nevada, Texas

Barcodes
0088483808526800884838099258
Units affected
108 systems (14 US, 1 Canada, 93 ROW)