Ge Healthcare Austria Gmbh & Co recalls GE Voluson IC9-RS intracavitary probes
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware of an issue with certain IC9-RS ultrasound probes (see Affected Products List) that can result in a double image artifact creating a ghost image with realistic features. An unrecognized artifact may lead to misdiagnosis.
- Products
-
- GE Voluson IC9-RS intracavitary probes
- Company
- Ge Healthcare Austria Gmbh & Co
- Where it was sold
-
worldwide
Nationwide
- Units affected
- 12004 units
- About this recall
- GE Voluson IC9-RS intracavitary probes, Model Number J48691PJ, Diagnostic ultrasound imaging and fluid flow analysis