Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ge Healthcare Austria Gmbh & Co recalls GE Voluson IC9-RS intracavitary probes

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware of an issue with certain IC9-RS ultrasound probes (see Affected Products List) that can result in a double image artifact creating a ghost image with realistic features. An unrecognized artifact may lead to misdiagnosis.
Products
  • GE Voluson IC9-RS intracavitary probes
Company
Ge Healthcare Austria Gmbh & Co
Where it was sold

worldwide

Nationwide

Units affected
12004 units
About this recall
GE Voluson IC9-RS intracavitary probes, Model Number J48691PJ, Diagnostic ultrasound imaging and fluid flow analysis