Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Howmedica Osteonics Corp. recalls SERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0515

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Products
  • SERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0515
Where it was sold

Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.

Nationwide

Barcodes
07613327034042
About this recall
SERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0515