Howmedica Osteonics Corp. recalls
Howmedica Osteonics Corp. is named on 15 recalls in Recall Relay's records since November 2023, from FDA (109 recalled products: FDA lists each product as its own recall); the most recent was September 16, 2026.
All 109 recalled products, newest first.
-
Class IIofficial
Howmedica Osteonics Corp. recalls TS Misc. Instruments TrayReprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
-
Class IIofficial
Howmedica Osteonics Corp. recalls TS Femoral Prep TrayReprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
-
Class IIofficial
Howmedica Osteonics Corp. recalls Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
-
Class IIofficial
Howmedica Osteonics Corp. recalls TS Universal Baseplate Finishing TrayReprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
-
Class IIofficial
Howmedica Osteonics Corp. recalls AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured fAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-413 Orthopaedic implanAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components mAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE BIPOLAR 40MM 6074-4022 SYSTEM ONE BIPOLAR 43MM 6074-4322 SYSAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M0Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG KNEE P/S 82-3-0318 SCORAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylenAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls OMNIFIT SERIES I ACET. INSERT 10¿ 22mm ID Catalog No. 2001-2246 Orthopaedic implant components manufactured frAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM 69-2244 ALL POLY CONSTRAINED INSERT 46MM 69-224Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls ABG I ACETABULAR INSERT HOODED 28mm ID X 48mm C-M079-2-801 ARC2F ACETABULAR INSERT 48/50/52mm OD x 28mm ID C-MAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 OrthopaeAffected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRI TS BASEPLATE SIZE 4. Part Number: 5521-B-400.Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300.Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
-
Class IIofficial
Howmedica Osteonics Corp. recalls MICS3 Angled Sagittal Saw Attachment; Part Number: 210490A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
-
Class IIofficial
Howmedica Osteonics Corp. recalls The Stryker CranialMask Tracker is a single-use device consisting of a flexible printed circuit board (PCB) wiWhen the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
-
Class IIofficial
Howmedica Osteonics Corp. recalls 1. EXETER V40 STEM 44MM NO 2a potential product mix where the package labeled as Catalog Number 0580-1-442, lot A00976, may contain Catalog Number 0580-1-352, lot G8754849, and vice versa. Patient labels may also be impacted.
-
Class IIofficial
Howmedica Osteonics Corp. recalls Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applicAnterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.
-
Class IIofficial
Howmedica Osteonics Corp. recalls MAKO Integrated Cutting System (MICS) HANDPIECEStryker has identified an issue with the characterization process that impacted the constant values assigned to the associated MICS Handpieces.
-
Class IIofficial
Howmedica Osteonics Corp. recalls Custom Made Device - GMRS EXTENSION Piece 40MM LEFTA patient specific, custom-made device was supplied for a surgical procedure that did not match the design proposal.
-
Class IIofficial
Howmedica Osteonics Corp. recalls UNITRAX Endoprosthesis Head Component - 42mm. Implantable component used in partial hip arthroplasty. CatalogMislabeled: the size on the package label of the UNITRAX¿ Endoprosthesis Head Component potentially may not match the device within the packaging. This product mix only involves the three (3) sizes (43mm, 42mm, and the 38mm)
-
Class IIofficial
Howmedica Osteonics Corp. recalls UNITRAX Endoprosthesis Head Component - 38mm. Implantable component used in partial hip arthroplasty. CatalogMislabeled: the size on the package label of the UNITRAX¿ Endoprosthesis Head Component potentially may not match the device within the packaging. This product mix only involves the three (3) sizes (43mm, 42mm, and the 38mm)
-
Class IIofficial
Howmedica Osteonics Corp. recalls UNITRAX Endoprosthesis Head Component - 43mm. Implantable component used in partial hip arthroplasty. CatalogMislabeled: the size on the package label of the UNITRAX¿ Endoprosthesis Head Component potentially may not match the device within the packaging. This product mix only involves the three (3) sizes (43mm, 42mm, and the 38mm)
-
Class IIofficial
Howmedica Osteonics Corp. recalls HRIS Storage & Sterilization Case and TrayThe design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
-
Class IIofficial
Howmedica Osteonics Corp. recalls Gray Revision Instrument System Accessory Tray/CaseThe design of the HRIS Storage & Sterilization Case and Tray P/N 6210-9-900 and the Gray Revision Instrument System Accessory Tray/Case P/N 6210-9-910 may lead to misuse of the Gray Revision Instruments including re-use and re-sterilization of the single-use, sterile-packaged instruments.
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII HEMI CLUSTER54E- Indicated for use in primary and revision patients at high risk of hip dislocationThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII HEMI CLUSTER44B- Indicated for use in primary and revision patients at high risk of hip dislocationThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII PSL CLUSTER50D- Indicated for use in primary and revision patients at high risk of hip dislocation CThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII HEMI CLUSTER58F- Indicated for use in primary and revision patients at high risk of hip dislocationThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII HEMI CLUSTER52E-Indicated for use in primary and revision patients at high risk of hip dislocation CThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII HEMI CLUSTER50D- Indicated for use in primary and revision patients at high risk of hip dislocationThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII PSL CLUSTER64H- Indicated for use in primary and revision patients at high risk of hip dislocation CThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII PSL CLUSTER52E- Indicated for use in primary and revision patients at high risk of hip dislocation CThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII HEMI CLUSTER48D-indicated for use in primary and revision patients at high risk of hip dislocation CThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII HEMI CLUSTER66H- Indicated for use in primary and revision patients at high risk of hip dislocationThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII PSL CLUSTER48D- Indicated for use in primary and revision patients at high risk of hip dislocation CThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls TRIDENTII PSL CLUSTER58F-Indicated for use in primary and revision patients at high risk of hip dislocation CaThe acetabular shell may have excessive deburring, resulting in a smooth surface on the edge of the shell
-
Class IIofficial
Howmedica Osteonics Corp. recalls NO 4 TRIATHLON TS PLUS TIB INS X3 POLY 16 MMPotential packaging breaches of inner blister and outer sterile blister.
-
Class IIofficial
Howmedica Osteonics Corp. recalls NO 4 TRIATHLON TS PLUS TIB INS X3 POLY 16 MMPotential packaging breaches of inner blister and outer sterile blister.
-
Class IIofficial
Howmedica Osteonics Corp. recalls X3 TRIATHLON CS INSERT NO 6 11 MMPotential packaging breaches of inner blister and outer sterile blister.
-
Class IIofficial
Howmedica Osteonics Corp. recalls X3 TRIATHLON CS INSERT NO 7 9 MMPotential packaging breaches of inner blister and outer sterile blister.
-
Class IIofficial
Howmedica Osteonics Corp. recalls NO 3 TRIATHLON TS PLUS TIB INS X3 POLY 11 MMPotential packaging breaches of inner blister and outer sterile blister.
-
Class IIofficial
Howmedica Osteonics Corp. recalls X3 TRIATHLON CS INSERT NO 3 10 MMPotential packaging breaches of inner blister and outer sterile blister.
-
Class IIofficial
Howmedica Osteonics Corp. recalls X3 TRIATHLON CS INSERT NO 6 10 MMPotential packaging breaches of inner blister and outer sterile blister.
-
Class IIofficial
Howmedica Osteonics Corp. recalls NO 4 TRIATHLON TS PLUS TIB INS X3 POLY 11 MMPotential packaging breaches of inner blister and outer sterile blister.
-
Class IIofficial
Howmedica Osteonics Corp. recalls HRIS ACET CUP CUT TIP 32X140 Part Number: 6210-5-200rker. single-useFailed to meet the acceptance criteria for the seal integrity and/or package integrity testing for the Sterile Barrier Outer Pouch and the Protective Barrier Inner Pouch, compromising product sterility
-
Class IIofficial
Howmedica Osteonics Corp. recalls HRIS ACET CUP CUT TIP 26X140 Part Number: 6210-5-100. single-useFailed to meet the acceptance criteria for the seal integrity and/or package integrity testing for the Sterile Barrier Outer Pouch and the Protective Barrier Inner Pouch, compromising product sterility
-
Class IIofficial
Howmedica Osteonics Corp. recalls DURATION A-P LIPPED TIB.INSERT-DURAC Intended for knee replacement Product Number: 6642-1-709Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls Scorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- -Intended for knee replacement Part Number: 72-16-0908Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3508Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-32JProduct manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls LARGE PRIMARY REV.TIB.INSERT- DURACON Intended for knee replacement Product Number: 6637-4-231Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TS TIB INSERT -Intended for knee replacment Part Number: 72-4-7516Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TS TIB INSERT. Intended for knee replacement Part Number: 72-4-7518Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls Scorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0908Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #5 -Intended for knee replacement Part Number: 73-0510Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0515Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0510Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls LEFT LARGE-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement Part Number: 6637-0-231Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls RGT.LARGE-PCA MOD.REV.- DUR.REV.INSERT Intended for knee replacement Product Number: 6637-0-831Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls Scorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0910Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3910Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3708Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls DUR PCA MTK REV INS RT Intended for knee replacement Product Number: 6728-2-709Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls RGT.LARGE-PCA MOD.REV.- DUR.REV.INSERT Intended for knee replacement Product Number: 6637-0-928Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7524Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TS TIB INSERT-Intended for knee replacement Part Number:72-4-7512Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls DURATION PLASTIC PATELLA-DURACON Intended for knee replacement Product Number: 6642-2-200Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #11 -Intended for knee replacement Part Number: 73-2110Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls P7 32MM 1O DEGREE +4MM INSERT Intended for knee replacement Part Number: 6302-6-307Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls LEFT X-LARGE-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement. Product Number: 6637-0-328Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls Scorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- Intended for knee replacement Part Number: 72-15-0324Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls LEFT LARGE-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement Part Number: 6637-0-228Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7521Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SERIES II TIBIAL BEAR INSERT-Intended for knee replacement Part Number: 3052-0524Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls RGT.LARGE-PCA MOD.REV.- DUR.REV.INSERT Intended for knee replacement. Product Number: 6637-0-931Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls P7 28MM 10 DEGREE +4MM INSERT Intended for knee replacement Part Number: 6302-6-107Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-28JProduct manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls DUR PCA MTK REV INS RT Intended for knee replacement Product Number: 6728-2-711Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-7514Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-413Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-416Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls TIBIAL BEARING INSERT SERIES P-S I ASSY Intended for knee replacement Product Number: 7291-0324Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3308Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls DUR PCA MTK REV INS LFT Intended for knee replacement Product Number: 6728-2-611Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-0910Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-0710Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls HOWMEDICA OSTEONICS TRIDENT 0 deg POLY INSERT Intended for knee replacement Part Number: 620-00-32JProduct manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls DUR PCA MTK REV INS LFT Intended for knee replacement Product Number: 6728-2-609Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO RECESSED PATELLA- Intended for knee replacement Part Number: 3044-0030Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls LEFT SMALL-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement Part Number: 6637-0-028Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TS TIB INSERT -Intended for knee replacement Part Number: 72-4-0321Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO RECESSED PATELLA-Intended for knee repalcment Part Number: 3044-0032Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO RECESSED PATELLA- Intended for knee replacement Part Number: 3044-0028Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO RECESSED PATELLA- Intended for Knee Replacement Part Number: 3044-0026Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls DURATION A-P LIPPED TIB.INSERT-DURAC Intended for knee replacement Product Number: 6642-1-911Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3110Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-2710Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-2910Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3710Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls LEFT SMALL-PCA MOD.REV.DUR.REV.INSERT Intended for knee replacement. Product Number: 6637-0-631Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #11 -Intended for knee replacement Part Number: 73-0110Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls PS LIPPED TIBIAL INSERT ASSY DURACON Intended for knee replacement Product Number: 6742-1-411Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls PRIMARY REV.TIB.INSERT-DURACON Intended for knee replacement. Product Number: 6637-4-031Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
-
Class IIofficial
Howmedica Osteonics Corp. recalls Scorpio-Flex Ttl Kn P-S Tib Brg Insrt Asy- -Intended for knee replacement Part Number: 72-16-0910Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.