Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Medical Systems China Co., Ltd. recalls GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU)

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
Products
  • GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU)
Where it was sold

Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lithuania, Malaysia, Mexico, Moldova, Morocco, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Thailand, Tunisia, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Venezuela and Vietnam.

Nationwide

Units affected
203 units
About this recall
GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S