GE Medical Systems China Co., Ltd. recalls GE Healthcare LOGIQ F R3 Ultrasound System
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.
- Products
-
- GE Healthcare LOGIQ F R3 Ultrasound System
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
00195278722430- Units affected
- 34 units
- About this recall
- GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system