Sin Sistema De Implante Nacional S.a recalls Implant Epikut S Plus CM 16-Degrees 4
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.
- Products
-
- Implant Epikut S Plus CM 16-Degrees 4
- Company
- Sin Sistema De Implante Nacional S.a
- Where it was sold
-
US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.
States: California, Colorado, Florida, Idaho, Illinois, Kentucky, New Jersey, New York, Ohio, Texas
- Barcodes
07899995273437- Units affected
- 123 (US 98, OUS 25)
- About this recall
- Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N