DICOM Grid, Inc. recalls Intelerad InteleShare software
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare viewer; Multiplanar reconstruction applied; Manual rotation applied, could result in inaccurate length measurement tool values that could compromise diagnostic accuracy, lead to misdiagnosis or inappropriate follow-up.
- Products
-
- Intelerad InteleShare software
- Company
- DICOM Grid, Inc.
- Where it was sold
-
US: AZ, TX, GA, NV, MO, MI, IL, OH, CT, OR, FL, NY, NM, MN, PA, AK, ND, NJ, VA, WA, NC, TN, WI, GA, AK, CA, SD, CO, MA, WV, MN, KS, MD, VT, IA, RI, AL, ID, LA, PR, NH, IN, OK, UT, NE, SC, AR, DE OUS: CA, NZ, AU, GB, BE, FR, IT, NO, SE
States: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kansas, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia
- Units affected
- 924
- About this recall
- Intelerad InteleShare software, with ProViewer component