Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM)

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.
Products
  • LifeShield Drug Library Management (DLM)
Where it was sold

US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.

States: California, Iowa, Kentucky, Louisiana, Maryland, Mississippi, New York, Pennsylvania, Tennessee, Virginia

Barcodes
00887709127227
Units affected
12 customers
About this recall
LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02