Draegerwerk Ag & Co. Kgaa recalls Atlan
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Internal backup battery may fail spontaneously while the Atlan is being operated without AC mains supply, resulting in unexpected shutdown of the device while running on batteries.
- Products
-
- Atlan
- Company
- Draegerwerk Ag & Co. Kgaa
- Where it was sold
-
No devices were distributed to end-users (customers) in the US. All devices are currently under the direct control of Draeger, Inc.
- Barcodes
04048675556176- Units affected
- 50 imported to US, not distributed to end users
- About this recall
- Atlan, Gas Machine for Anesthesia or Analgesia, Catalog numbers 8621500 (Atlan A350) and 8621600 (Atlan A350 XL)