Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Draegerwerk Ag & Co. Kgaa recalls Atlan

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Internal backup battery may fail spontaneously while the Atlan is being operated without AC mains supply, resulting in unexpected shutdown of the device while running on batteries.
Products
  • Atlan
Company
Draegerwerk Ag & Co. Kgaa
Where it was sold

No devices were distributed to end-users (customers) in the US. All devices are currently under the direct control of Draeger, Inc.

Barcodes
04048675556176
Units affected
50 imported to US, not distributed to end users
About this recall
Atlan, Gas Machine for Anesthesia or Analgesia, Catalog numbers 8621500 (Atlan A350) and 8621600 (Atlan A350 XL)