Randox Laboratories Ltd. recalls RX Series Copper (Cu) Assay Ref. Number CU2340
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Randox Laboratories have released an update to the Carryover Avoidance Technical Bulletin to detail that Copper, should not be run directly after Total Protein, on the RX series instruments testing order. If copper assay is run after total protein it could lead to falsely elevated copper results.
- Products
-
- RX Series Copper (Cu) Assay Ref. Number CU2340
- Company
- Randox Laboratories Ltd.
- Where it was sold
-
US Nationwide distribution in the states of SC and OK.
States: Oklahoma, South Carolina
- Units affected
- 3,053 kits (2 US, 3,051 OUS)
- About this recall
- RX Series Copper (Cu) Assay Ref. Number CU2340